Evaluation of the frequency stability of a VCSEL locked to a micro-fabricated Rubidium vapour cell

Auteur(s)
Di Francesco, Joab F.
Gruet, Florian 
Institut de physique 
Schori, Christian 
Institut de physique 
Affolderbach, Christoph 
Institut de physique 
Matthey-De-L'Endroit, Renaud 
Institut de physique 
Mileti, Gaetano 
Institut de physique 
Salvadé, Y.
de Rooij, Nicolaas F.
Petremand, Y.
Date de parution
2010
In
Semiconductor Lasers and Laser Dynamics IV, Society of Photo-optical Instrumentation Engineers (SPIE), 2010/7720/77201T/1-9
Mots-clés
  • Vertical cavity surface emitting laser (VCSEL)
  • micro-fabricated absorption cell
  • Rubidium
  • laser frequency stabilization
  • Allan deviation
  • spectroscopy
  • Vertical cavity surfa...

  • micro-fabricated abso...

  • Rubidium

  • laser frequency stabi...

  • Allan deviation

  • spectroscopy

Résumé
We present our evaluation of a compact laser system made of a 795 nm VCSEL locked to the Rubidium absorption line of a micro-fabricated absorption cell. The spectrum of the VCSEL was characterised, including its RIN, FM noise and line-width. We optimised the signal-to-noise ratio and determined the frequency shifts versus the cell temperature and the incident optical power. The frequency stability of the laser (Allan deviation) was measured using a high-resolution wavemeter and an ECDL-based reference. Our results show that a fractional instability of ≥ <sup>10-9</sup> may be reached at any timescale between 1 and 100'000 s. The MEMS cell was realised by dispensing the Rubidium in a glass-Silicon preform which was then, sealed by anodic bonding. The overall thickness of the reference cell is 1.5 mm. No buffer gas was added. The potential applications of this compact and low-consumption system range from optical interferometers to basic laser spectroscopy. It is particularly attractive for mobile and space instruments where stable and accurate wavelength references are needed.
Identifiants
https://libra.unine.ch/handle/123456789/12190
_
10.1117/12.854147
Type de publication
journal article
Dossier(s) à télécharger
 main article: Di_Francesco_J._-_Evaluation_of_the_frequency_stability_of_VCSEL_locked_20101217.pdf (1.08 MB)